Speciation of Manganese Using a Pneumatic Flow Injection Analysis-Tandem Spectrometer System

Authors: MEISSAM NOROOZIFAR, MOZHGAN KHORASANI-MOTLAGH, REZA AKBARI

Abstract: A pneumatic flow injection-tandem spectrometer system, without a delivery pump, was used for the speciation of manganese (Mn). In this system the suction force of the pneumatic nebulizer of a flame atomic absorption spectrometer (FAAS) was used for solution delivery through the manifold. Mn(VII) and total Mn (Mn(VII) and Mn(II)) concentrations were determined using a UV-Vis spectrometer and FAAS, respectively. The Mn(II) concentration was determined by the difference between the two. The calibration curves were linear up to 15.00 mg L^{-1} for Mn(VII) and total Mn, with a detection limit of 0.08 mg L^{-1} and 0.05 mg L^{-1} for Mn(VII) and Mn, respectively. The mid-range precision and accuracy were < 1.89% and \pm 2.50% for the 2 species, respectively, at a sampling rate of 80 samplings h^{-1}. This system was used for the determination of Mn(VII) and Mn(II) in spiked and natural water, as well as industrial water.

Keywords: Pneumatic flow injection analysis, tandem spectrometer, manganese, speciation, UV-Vis spectrometer, flame atomic absorption spectrometer

Full Text: PDF